PDS_VERSION_ID = PDS3                                                         
                                                                              
RECORD_TYPE  = STREAM                                                         
INSTRUMENT_HOST_NAME = "ROSETTA-ORBITER"                                      
DATA_SET_ID = "RO-C-RPCICA-3-ESC3-CALIB-V1.0"                                 
PRODUCT_CREATION_TIME = 2020-10-30T08:00:23                                   
PRODUCT_ID = "CALIBRATION.TXT"                                                
                                                                              
OBJECT     = TEXT                                                             
 PUBLICATION_DATE   = 2020-10-30                                              
 INTERCHANGE_FORMAT = ASCII                                                   
 NOTE               = "Description of calibration files"                      
END_OBJECT = TEXT                                                             
END                                                                           
***** File CALIBRATION.TXT                                                    
                                                                              
Calibration files for RPC-ICA data                                            
                                                                              
The level 3 data is identical to the level 2 edited raw data                  
except that calibration constants have been used to turn the                  
data into physics units, that of differential                                 
flux [particles / cm^2 / s / sr / eV].                                        
                                                                              
The procedure to turn level 2 data into level 3 data is described             
in general terms in the EAICD, section 2.4.1.                                 
                                                                              
In short, data is divided into light or heavy ions using the mass             
masks provided in the directory CALIB/MASS_MASK. Then the geometric           
factor found in CALIB/GFACTOR is applied according to the formula             
given in the EAICD. Physical elevation angles corresponding to                
different elevation index levels are found in the tables in                   
CALIB/ELEVATION and the energy tables providing energy for each               
energy index level are found in CALIB/ENERGY.                                 
                                                                              
Different tables for MASS_MASK and GFACTOR exist for different                
post-acceleration settings and different software versions and                
for two ion mass ranges, light (approximately 1 to 4 amu) and                 
heavier ions. For elevation and energy there is one table for                 
each software version.                                                        
                                                                              